DwyerOmegaHOBO-MX1104
MX1104
- 판매 가격
- 가격 문의
수량
재고: 0개국내 재고가 없어도 주문 가능합니다. 주문 시 해외 재고를 확인합니다.
무료 배송 10만원 이상
납기: 13 weeks리드타임은 상황에 따라 달라질 수 있습니다.
연구비카드 사용가능
실시간 문의하기제품 사양
Brand:OMEGA Engineering
Accuracy Detail:Temperature Sensor Accuracy: ± 0.20°C from 0° to 50°C (± 0.36°F from 32° to 122°F), Current Accuracy: ± 0.001 mA ± 0.2% of reading, Voltage Accuracy: 0 to 24 V: ± 1.0 mV ± 0.3% of reading, 0 to 2.5 V: ± 0.1 mV ± 0.1% of reading, 0 to 5 V: ± 0.2 mV ± 0.3% of reading, 0 to 10 V: ± 0.4 mV ± 0.3% of reading, RH: ± 2.5% from 10% to 90% (typical) to a maximum of ± 3.5% including hysteresis at 25°C (77°F); below 10% and above 90% ± 5% typical
Approvals:CE
Battery Size and Type:Alkaline AAA
Battery life:1 yr, typical with reading interval of 1 min
Configuration:HOBOmobile® app Android, iOS, iPadOS
Current Accuracy:± 0.001 mA
Current Measurement Range:0 to 20.1 mA
Current Resolution:Temperature: 0.002°C, Relative Humidity: 0.01%, Voltage: 384 µV, Current: 0.3 µA
Data Interface:Bluetooth
Display:LCD
Display Note:LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Enclosure Ratings:IP54
Form Factor Note:Use the four magnets on the back of the logger to mount it to a magnetic surface or insert the tie wraps through the mounting loops to attach it to other surfaces
Form factor:Magnetic Surface or Insert the tie wraps
Height:1.15 in
Input Type:Internal, External
Length:2.13 in
Light Accuracy:± 10%
Light Measurement Range:0 to 167,731 lux
Logger with Cable Type:"Current: SD-MA-420 or CABLE-4-20mA Voltage: 0 to 24 V: SD-VOLT-24 or CABLE-ADAP24 0 to 2.5 V: SD-VOLT-2.5 or CABLE-2.5-STEREO 0 to 5 V: SD-VOLT-05 or CABLE-ADAP5 0 to 10 V: SD-VOLT-10 or CABLE-ADAP10"
Logging Modes:burst, Fixed interval
Measurement Type:Current, Light Intensity, Relative Humidity, Temperature, Voltage
Memory:4 MB
Memory Modes:stop when full, Wrap when full
Memory Note:Approximately 4 to 15 minutes depending on the mobile device; may take longer the further the device is from the logger
Number of Measurements:4
Operating Temperature, Max:70 °C
Operating Temperature, Min:-20 °C
RH Measurement Range:0 to 100% RH
RH Resolution:0.0001
RH Sensor Drift:<1% Per Year
Radio Power:1 mW
Relative Humidity Accuracy:± 2.5% RH
Resistance Resolution:Temperature: 0.002°C, Relative Humidity: 0.01%, Voltage: 384 µV, Current: 0.3 µA
Resolution Note:Temperature Sensor Resolution: 0.002°C at 25°C
Response Time:Temperature: 11 Min, RH: 30 sec
Response Time Note:Temperature: 11 Minutes in air moving 1 m/s, RH: 30 seconds to 90% in airflow of 1 m/s
Restart Mode:Push button
Samples,Max:~1.9M Samples, Interval Selectable 1 sec to 18hrs
Start Modes:At set date and time, Button Push, Immediate Start, Next Interval
Stop Modes:Button Push, When Memory is Full, At set date and time, At set Duration
Supply Power Type:Battery
Temperature Accuracy:± 0.20°C
Temperature Resolution:0.002 °C
Temperature Sensor Drift:< 0.1°C Per Year
Time Accuracy:± 1 minute/month at 25°C
Transmission Range:30.5 m
Voltage Accuracy:± 1.0 mV
Voltage Measurement Range:0 to 2.5 V, 0 to 5.0 V. 0 to 10 V, 0 to 24 V
Voltage Resolution:384 µV
Weight:4.34 oz
Width:4.44 in