DwyerOmegaCT5100-Series
CT5102
1 or 2 Channel, 10" Circular Chart Recorders
- 가격
- 4,507,940 원
- VAT 10%
- 450,794 원
- 총 금액
- 4,958,734 원
수량
재고: 0개국내 재고가 없어도 주문 가능합니다. 주문 시 해외 재고를 확인합니다.
무료 배송 10만원 이상
납기: 18 weeks리드타임은 상황에 따라 달라질 수 있습니다.
연구비카드 사용가능
실시간 문의하기제품 사양
Accuracy:± 0.5 %
Accuracy Detail:Chart Recording: 0.5% of chart span reference accuracy, Chart Rotation: ±0.5% of rotation time, assuming all backlash removed, Chart Rotation: ±0.5%, Chart Speed: ±1%
Approvals Note:Safety: UL Approved for USA:UL 1092, UL 916 and QUXY, pending UL Certified for Canada:CSA Spec 142, pending Limit/Device: FM, pending
Chart Diameter:254 mm
Chart Note:Chart Span: Bottom and top of span, -9999 to 9999 units; Chart Drive: DC Stepper Motor; Chart Drive: Stepper Motor
CMRR:>120 dB @ 50/60 Hz, 260 Vac max
Cold Junction Compensation:0.04°/°C deviation from 25°C
Conduit Entrance:3 Openings on the Right Side
Current Measurement Range:0 to 20 mA
Cutout Dimensions:322.58 x 322.58 mm
Data protection:EEPROM Storage
Display Note:4-digit, 14 mm H (0.56"), red, 7-segment, LED display, Display Modes: Normal: process value(s) or blank Keypad: 3 keys for programming and unit operation
Display Type:LED
Enclosure Ratings:IP54, NEMA Type 3
Height:355.6 mm
Impedance Input:>100 MΩ for T/C and mV inputs, 100 kΩ for 5 V inputs, 4.7 Ω for µA inputs
Input Note:Input Correction: Offset adjustment: 999 to 999 units;Input Isolation: 350 Vac, 500 Vdc
LED Indicators:5 Red LED Alarm Status Indicators, 1 green LED pen 2 indicator
Length:96.5 mm
Materials:Noryl, Acrylic
Measurement Range Note:Thermocouple: J: 0 to 760°C, K: 0 to 1360°C, T: -200 to 400°C, R: 200 to 1650°C, S: 200 to 1650°C RTD: -140 to 400°C Current: 0 to 20 mA, 4 to 20 mA Voltage: 0 to 25 mV, 0 to 50 mV, 10 to 50 mV, 0 to 5 V, 1 to 5 V
Measurement Type:Thermocouple, RTD, Current, Voltage
Mounting:Panel Mount or Wall Mounting
NMRR:85 db @ 60 Hz or greater
Operating Environment:Reference Operating Condition: 25 °C
Operating Temperature, Max:50 °C
Operating Temperature, Min:0 °C
Panel Depth:64 mm
Panel Protrusion:33 mm
Pen Type:Disposable Fiber tip
Power Consumption:18 VA
Relative Humidity:10 to 90% RH, non-condensing
RTD Type:100 Ω Pt sensor; α= 0.00385
Sampling Rate:1 Scan per second for Non-RTD inputs; 1 Scan per 1.2 seconds for RTD inputs
Sensor Input Configuration:150 µA
Series ID:CT5100-Series
Standard Limits of Error:Error Limit: Cold-Junction Compensation Error: ±0.2°C @ 25°C, Linearization Error: T/Cs: ±0.25°C typical, ±0.5°C worst case , RTDs: ±0.1°C typical, ±0.3°C worst case, Ambient Temperature Error: ±0.01% of span per °C deviation from 25°C; Measurement Error: Type J, K, T, R, S and RTD: ±0.25% of span ±1 degree; mA, mV and Vdc: ±0.25% of scaled span plus 1 leastsignificant digit
Storage Temperature, Max:65 °C
Storage Temperature, Min:-40 °C
Supply Power Note:Line Voltage: 90/264 Vac, 50/60 Hz; optional: 20 to 50 Vac, 50/60 Hz, or 22 to 65 Vdc
Supply Power Type:AC
Supply Voltage:90 to 264 Vac
Switch Rating:5 A @ 115 Vac
Switch Rating Note:Relays: SPDT; contacts rated 5 A resistive @ 115 Vac,2.5 A resistive @ 230 Vac, 1⁄8 hp @ 230 Vac (single phase),250 VA @ 115/230 Vac
Switch Type:SPDT NO/NC, Selectable
Temperature Measurement Range:J: 0 to 760°C, K: 0 to 1360°C, T: -200 to 400°C, R: 200 to 1650°C, S: 200 to 1650°C, RTD: -140 to 400°C
Thermocouple Type:J, K,T, R, S
Voltage Measurement Range:0 to 25 mV, 0 to 50 mV, 10 to 50 mV, 0 to 5 V, 1 to 5 V
Weight:8.8 kg
Width:355.6 mm
Wire Code Option:2- or 3- Wire