DwyerOmegaCN8200-Series
CN8202-R1-R2-AL1
1/16 Din Universal Temperature & Process Controllers with 8 Segment Ramp/Soak Capability
- 판매 가격
- 가격 문의
수량
재고: 0개국내 재고가 없어도 주문 가능합니다. 주문 시 해외 재고를 확인합니다.
무료 배송 10만원 이상
납기: 10 weeks리드타임은 상황에 따라 달라질 수 있습니다.
연구비카드 사용가능
실시간 문의하기제품 사양
Brand:OMEGA Engineering
Accuracy:± 0.2 %
Accuracy Detail:± 0.2% FS, ± 1 digit
CMRR:100 dB
Cold Junction Compensation:0.05°C/°C ambient
Contacts:Twin bifurcated
Control Characteristics:Set point Limits: Limited to configured range for thermocouple and RTD; limited to scaled range; Alarms: Adjustable for high/low; selectable process or deviation; Rate (Derivative): 0 to 2400 seconds; Reset (Integral): 0 to 9600 seconds; Cycle Time: 0.2 to 120 seconds; Proportional Band: 1 to span of sensor; Dead band: Negative span to positive span of sensor; Hysteresis: 1 to span of sensor; Autotune Damping: Adjustable (low, normal, or high); Ramp to Set point: 1 to 9999 minutes; Autotune: Operator-initiated from front panel; Manual Control: Operator-initiated from front panel
Control Output:SPDT Relay, 2 NO SPST Relay
Cutout Dimensions:45 x 45 mm
Data Interface:None
Depth Behind Panel:100 mm
Display Size:9.2 mm (.36 in)
Display Type:4 Digit LED
Display Type Note:Dual LED—4-digit, orange: process display; green: menu/parameter display; 9.2 mm (0.36")
Electrical Connection:Screw Terminal
Enclosure Ratings:IP65, NEMA Type 4X
Events/Alarms:1 to 8
Form factor:1/16 DIN
Hardware and Software Requirements, Min:CN8-SW (Optional Software): Minimum Hardware and Software Requirements: IBM PC or 100% compatible, Windows 95/98/NT; RS485 interface or RS232 to RS485 converter; Software Compatibility: CN8200 Series controllers; Software Capability: Supports up to 254 CN8200 Series controllers
Height:53.3 mm
Input Type:Thermocouple, RTD, mA, Voltage
Input Type Note:Input Type: See input table on next page Digital Input: For remote setpoint, remote standby or ramp/soak run and hold Thermocouple Lead Resistance: 100 Ω maximum for rated accuracy Decimal Position: Selectable
Length:8.21 mm
Loops:0 to 99
Measurement Type:Current, Temperature, Voltage
Measuring Range:Type J (Iron-Constantan): -200 to 760°C, Type K (CHROMEGA®-ALOMEGA®): -270 to 1354°C, Type T (Copper-Constantan): -270 to 400°C, Type N (OMEGALLOY®) : -268 to 1300°C, Type R (Pt/13%Rh-Pt) : -50 to 1768°C, Type S (Pt/10%Rh-Pt) : -50 to 1768°C, Type B (Pt/30%Rh-Pt/6%Rh) : 0 to 1820°C, Type C (W/5%Re-W/26%Re) : 0 to 2315°C, Type E (CHROMEGA®-Constantan) : -150 to 1000°C, NNM 18% molybdenum vs nickel -06% cobalt : 0 to 1410°, Platinel II : -100 to 1232°C, RTD ((3-wire) 100 Ω Pt) : -200 to 850°C, RTD ((3-wire) 100 Ω Pt) : -199.9 to 375.0°C, 0 to 1V : Scalable (-1999 to 9999) selectable 1 to 5V : Scalable (-1999 to 9999) selectable 0 to 5V : Scalable (-1999 to 9999) selectable 0 to 10V : Scalable (-1999 to 9999) selectable 10 to 50 mV : Scalable (-1999 to 9999) selectable 0 to 50 mV : Scalable (-1999 to 9999) selectable 0 to 10 mV : Scalable (-1999 to 9999) selectable 0 to 100 mV : Scalable (-1999 to 9999) selectable 4 to 20 mA : Scalable (-1999 to 9999) selectable 0 to 20 mA : Scalable (-1999 to 9999) selectable
Memory Backup:Solid State
Number of Control Outputs:2
Number of Inputs:1
Operating Temperature, Max:55 °C
Operating Temperature, Min:0 °C
Output Detail:Relay, Relay, Mechanical Relay;Output #1: Reverse- or direct-acting, configured from menu Output #2: Reverse- or direct-acting, configured from menu Mechanical Relay: Rated 5 A @120 Vac, 3 A @ 240 Vac, normally open (NO), normally closed (NC) (output 1); rated 5 A @120 Vac, 3 A @ 240 Vac, NO (output 2) Current: 4 to 20 mA, 500 Ω maximum (suffix F1, F2); 4 to 20 mA, 1000 Ω maximum (suffix FH1, FH2) Voltage: 20 Vdc pulse Solid State Relay: SSR, 120/240 Vac, zero voltage switched, rated 1 A continuous, 10 A surge @ 25°C (77°F) Transducer Power Supply: 5, 10, 12, 15 Vdc ± 10% ;Alarm Output:Mechanical relay rated 5 A @ 120 Vac, 3 A @ 240 Vac, NO; optically isolated SSR rated 1 A, 120/240 Vac @ 25°C (77°C); DC alarms, 24 Vdc
Output Signal:None
Ramp Setpoint:1 to 9999 minutes
Ramp and Soak:8 Ramp & Soak
Ramp/Soak Time:0 to 9999 minutes
Relative Humidity:90% RH, non-condensing
Remote Setpoint:N/A
Repeatability:± 1 count
Sampling Rate:10 Hz (10 updates/second)
Series Mode Rejection:>70 dB
Setpoint Resolution:1 count/0.1 count
Supply Power Note:100 to 250V, 50/60 Hz (single-phase); 120 to 250 Vdc, 24 Vac/24 Vdc (optional)
Supply Power Type:AC
Supply Voltage:100 to 250 Vac
Temperature Stability:5 μV/°C Max
Weight:226 g
Width:53.3 mm